PanAn Testing & Engineering Co.,Ltd.
Welcome to PanAn Testing & Engineering Co.,Ltd.
Products
Location position:Products
>
<
TO10 Thermal Oxidation Stability Test Rig AD systems

Category: Others

Method Standard: ASTM D3241, IP 323, ISO 6249

Application: Oil Refining

Brand: AD systems

Hotline:
021 - 63175588
Products description Products characteristic technical specification

For the development of this innovative instrument, all the parameters influencing the accuracy of

the D3241 / IP323 test method have been identified. For
each of these parameters an innovative and reliable
solution has been developed. This very successful
apparatus is unmatched in the market.
The preparation of the test is simplified. No tools are
needed for assembly and the heating tube mounting.
An extremely intuitive teaching software makes it
possible to very quickly initiate a test and if necessary to
follow step by step its progress.

Dual 5 mL syringe fuel pumping system, perfect fuel flow control, no pulse

Automatic fuel aeration control

Didactic and intuitive man-machine interface
Independent bus bar temperature control, no cooling liquid
DR10 – ITR connectivity for automatic result storage

Test method:ASTM D3241, IP 323, ISO 6249,Up to 21 programmable test methods

Test Temperature: Programmable from 100°C up to 380°C
Differential Pressure:0 – 750 mmHg (automatically bypassed at +750 mmHg)

Test time Range:Programmable from 4 to 600 minutes

Fuel Aeration: Programmable time, automatic air flow control

Aeration Flow rate:Programmable or according to test method, 1.5 mL / min

Adjustable flow rate 0.1ml/min~9.99ml/min

Heater tube temperature measurement: Thermocouple Type K, class 1
Fuel flow: fuel flow rate 3.45MPa (500psi±40psi); 3ml/min±0.2ml/min
Accuracy: ±1% No flow pulse or peak

Bus Bar Temperature control:Independent and programmable, each bus bar temperature can be adjusted,No liquid cooling circulation,Peltier modules + heat pipe technology


Heater tube section assembling: Special gauge to quickly and perfectly position the heater tube

Diagnostic and Service:Dedicated service menu with a flow chart layout with ability to click on each symbolic element to check its operation



Results storage: results database, limited to the capacity of the external device
LAN connection: Ethernet port RJ45
Printer output: USB (printer is optional)
Data output: USB (2), Ethernet
Dimensions (mm): 440 x 600 x 670 (17" x 23" x 26")
Weight: 60 kg (133 lbs)
Electrical: 100 to 240V - 8A - 50/60 Hz
Operating temperature: +10 to +35°C