PanAn Testing & Engineering Co.,Ltd.
Welcome to PanAn Testing & Engineering Co.,Ltd.
Products
Location position:Products
>
<
TE-3000 Sub-Zero Constant Temperature Bath CANNON

Category: Viscosity Measurement

Method Standard: ASTM D445,D446,GB/T 265, ISO 3104,3105, IP 71

Application: Lubricants

Brand: CANNON

Hotline:
021 - 63175588
Products description Products characteristic technical specification
The microprocessor controlled TE-3000 series constant temperature bath, with integrated thermoelectric cooling, accommodates viscometer tubes requiring a bath depth up to 12” and provides reliable temperature control from –30 °C to 30 °C. The unit accommodates up to 2 simultaneous viscosity determinations and offers multi-unit networking capability.

Precise & accurate temperature control

- Temperature range: 30 °C to 30 °C ± 0.01 °C

- Meets ASTM D445 temperature sensitivity requirements for low temperature kinematic viscosity

Microprocessor-controlled temperature regulation & networking capabilities

- Desired bath temperature and calibration temperatures are digitally entered on the front panel keypad

- Integrated microprocessor automatically regulates bath temperature

- Both target and actual bath temperatures are monitored by an RTD temperature probe and continuously displayed in °C or °F on the LCD screen

- RS-232 and RS-485 communication ports provide data acquisition and multi-unit networking capabilities

Compact, user-friendly design

- Environmentally-friendly, integrated, thermoelectric air cooling requires no external refrigeration or air/water heat exchangers

- Two glass viscometer ports (each 2in diameter) and one thermometer port (0.4” in diameter) allow up to 2 simultaneous

viscosity determinations

- Baffl e within the aluminum bath vessel facilitates circulation for uniform bath temperature and provides a plain viewing

background

- Frost-free viewing window and internal illumination provide excellent visibility within the bath

Safe operation
- Bath thermistor senses over-temperature fault and removes power
- Thermistor disconnection results in removal of bath heater power
Reliable performance
- Well tested CANNON bath platform off ers reliability and dependable support
- The heavy duty aluminum frame off ers a corrosion-resistant epoxy finish