Category: Viscosity Measurement
Method Standard: ASTM D445,D446,GB/T 265, ISO 3104,3105, IP 71
Application: Lubricants
Brand: CANNON
The CT-500 series constant temperature bath accommodates viscometer tubes requiring a bath depth up to 12” and provides reliable temperature control from 20 °C to 100 °C . The unit accommodates up to 7 simultaneous viscosity determinations.
Precise & accurate temperature control
- Temperature range: 20 °C to 100 °C ± 0.01 °C*
- Temperature selection is variable between 20 °C and 100 °C or selectable at ten pre-set temperatures (20 °C, 25 °C, 30° C, 40 °C, 50 °C, 60 °C, 70 °C, 80 °C, 90 °C, and 100 °C)
- Meets ASTM D445 temperature sensitivity requirements for kinematic viscosity determination
- Motor driven stirrer ensures a uniform bath temperature
Safe operation
- Bath thermistor senses over- temperature fault and removes power
- Thermistor disconnection results in removal of bath heater power
- Liquid level sensor prevents operation if bath liquid is not at a safe operating level
Efficient, user-friendly design
- Seven glass viscometer ports (each 2” in diameter) and two thermometer ports (each 0.4” in diameter) allow up to 7 simultaneous viscosity determinations**
- Two electric heating elements quickly heat bath media to the selected temperature
- Stainless steel baffle provides a plain reflective viewing background
- Convenient drawer provides easy access to electrical controls