PanAn Testing & Engineering Co.,Ltd.
Welcome to PanAn Testing & Engineering Co.,Ltd.
Products
Location position:Products
>
<
CT-1000 Extended Range Constant Temperature Bath CANNON

Category: Viscosity Measurement

Method Standard: ASTM D445,D446,GB/T 265, ISO 3104,3105, IP 71

Application: Lubricants

Brand: CANNON

Hotline:
021 - 63175588
Products description Products characteristic technical specification

ASTM D445, ISO3104, GB/T 265

Temperature controlled20~ 100℃±0.01/over 100℃ ±0.03

The CT-1000 series constant temperature bath accommodates viscometer tubes requiring a bath depth up to 12and provides reliable temperature control from 20 °C to 150 °C . The unit accommodates up to 7 simultaneous viscosity determinations.



Precise & accurate temperature control

-- Temperature range: 20 °C to 150 °C*
-- Temperature accuracy: ± 0.01 °C from 20 °C to 100 °C; ± 0.03 °C from 100 °C to 150 °C
-- Temperature selection is variable between 20 °C and 150 °C or selectable at ten pre-set temperatures (20 °C, 25 °C,

30° C, 40 °C, 50 °C, 60 °C, 80 °C, 100 °C, 135 °C and 150 °C)
-- Meets ASTM D445 temperature sensitivity requirements for kinematic viscosity determination
-- Motor driven stirrer ensures a uniform bath temperature

Safe operation
-- Bath thermistor senses overtemperature fault and removes power
-- Thermistor disconnection results in removal of bath heater power
-- Liquid level sensor prevents operation if bath liquid is not at a safe operating level

Efficient, user-friendly design
-- Seven glass viscometer ports (each 2” in diameter) and two thermometer ports (each 0.4” in diameter) allow up to 7 simultaneous viscosity determinations**
-- Two electric heating elements quickly heat bath media to the selected temperature